ABB TB82PH Istruzioni per gli Interventi Pagina 7

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 6
I-E67-TB82FF-2A February 13, 2002
7
INSTALLATION
Besides the normal precautions for storage and
handling of electronic equipment, the transmitter
contains static sensitive devices. Since
semiconductors can be damaged by the direct
discharge of static electricity, avoid contact with
terminal block conductors and electronic components
on the circuit board.
When mounting the transmitter, choose a location that
has ample clearance for the removal of the front bezel
and rear cover. The location should provide easy
access for maintenance procedures and not be in a
highly corrosive environment. Excessive mechanical
vibrations and shocks as well as relay and power
switches should not be in the immediate area. Signal
wiring should not be placed in conduit or open trays
that contain power wiring for heavy electrical
equipment. Field wiring should meet wiring practices
appropriate for fieldbus devices (See Supplementary
Documentation).
The transmitter can be pipe, hinge, wall or panel
mounted. Use the appropriate figure below to assist
in the mechanical installation of the transmitter.
3/8” NUT
4TB4711-0020
(4 TYP)
3/8”
FLATWASHER
4TB4710-0028
(8 TYP)
INSTRUMENT
MOUNT
BRACKET
4TB5008-0071
3/8” X 5/8”
BOLT
4TB4704-0119
(4TYP)
3/8”
LOCKWASHER
4TB4710-0022
(8 TYP)
3/8” X 3/4”
BOLT
4TB4704-0086
(4 TYP)
5/16” NUT
4TB4711-0013
(4 TYP)
5/16”
FLATWASHER
4TB4710-0025
(4 TYP)
5/16” U-BOLT
4TB4704-0096
(2 TYP)
PIPE MOUNT
BRACKET
4TB5008-0022
5/16”
LOCKWASHER
4TB4710-0023
(4 TYP)
Figure 5 – Pipe Mounting Detail
FRONT VIEW
3/8”
FLATWASHER
4TB4710-0028
(4 TYP)
3/8” X 5/8”
BOLT
4TB4704-0119
(4TYP)
INSTRUMENT
MOUNT
BRACKET
4TB5008-0071
TOP VIEW
FASTENERS FOR
WALL (SUPPLIED
BY OTHERS)
S.S. HINGE
4TB5010-0005
3/8” X 3/4”
BOLT
4TB4704-0086
(8 TYP)
3/8” NUT
4TB4711-0020
(8 TYP)
3/8”
FLATWASHER
4TB4710-0028
(8 TYP)
3/8”
LOCKWASHER
4TB4710-0022
(12 TYP)
“L” BRACKET
4TB5008-0073
Figure 6 – Hinge Mounting Detail
3/8”
LOCKWASHER
4TB4710-0022
(4 TYP)
3/8” X 5/8”
BOLT
4TB4704-0119
(4 TYP)
FASTENERS FOR WALL
(SUPPLIED BY OTHERS)
3/8”
FLATWASHER
4TB4710-0028
(4 TYP)
INSTRUMENT
MOUNT
BRACKET
4TB5008-0071
Figure 7 – Wall Mounting Detail
REMOVE EXTERNAL GROUND SCREW AND HARDWARE BEFORE INSTALLING
PANEL MOUNT BRACKET ASSEMBLY IF NECESSARY.
INSERT INSTRUMENT INTO PANEL CUT-OUT AND INSTALL PANEL MOUNT
BRACKET ASSEMBLIES.
TIGHTEN PANEL MOUNT SCREWS ON PANEL MOUNT BRACKET ASSEMBLY UNTIL
GASKET SEATS AGAINST WALL. DO NOT OVER-TIGHTEN PANEL MOUNT SCREWS
OR DAMAGE TO THE PANEL MOUNT BRACKET ASSEMBLY AND/OR PANEL MAY OCCUR.
PANEL CUT-OUT
5.33
135.4
+0.05
-0.03
+1.3
-0.8
0.47
11.9
5.33
+0.05
-0.03
135.4
+1.3
-0.8
11.9
0.47
1
2
3
1
3
3
2
3
3
2
GASKET
WALL
EXTERNAL
GROUND
SCREW
REAR VIEW
3/8”
MAX
WALL
REAR COVER
REMOVED
Figure 8 – Panel Mounting Detail
Warning.
For installation in Hazardous Areas, i.e. areas
with danger of fire and/or explosion,
irrespective of the protection mode used, the
installation must carried out in accordance with
local authority regulations. Ensure also that the
temperature of the transmitter does not exceed
the value indicated in the Safety Marking plate.